The cyclical system water exchange ensures that the heat transfer medium in the temperature control unit does not change. The water fed in via the cooling water inlet or system water inlet remains in the heat transfer circuit during operation. Cyclical system water exchange is only recommended when using treated water. If the cyclic system water exchange act switched on, rinsing takes place at the set time interval.
ATTENTION!
Damage due to incorrect setting If the water quality is poor, the cyclical system water exchange can lead to fouling and scaling. That is why:
Only switch on the cyclic system water exchange when the water has been well treated.
Use the slider (, ) to switch the function on or off.
NOTE!
The cyclic system water exchange is switched off as standard.
A value set too low can lead to fluctuations in temperature control.
For well-treated water, a rinse interval of is recommended. 30 min With this setting and a rinsing time of, the medium in the circuit 0,5 s is replaced in 1–2 days under normal operating conditions.
A value set too high can lead to fluctuations in temperature control.
A rinsing time of 0,5 s is recommended.
Datenschutz - Data protection - Protection de données
Wir setzen Cookies ein, um Ihnen die Nutzung unserer Webseite zu erleichtern. Weitere Informationen zu Cookies finden Sie in unserer Datenschutzerklärung. We use cookies to help you use our website. More information about cookies can be found in our privacy policy. Notre site utilise des cookies afin de vous garantir un service optimal. En utilisant ce site web, vous acceptez les présentes conditions. Vous trouverez privacy policy de plus amples informations. Privacy Policy