Download page Menu structure Thermo-6.
Menu structure Thermo-6
Main menu | Submenu 1 | Submenu 2 | Submenu 3 | Submenu 4 | Profile | Operat. release | Default | Units | Reset custom | Reset system | Reset complete |
---|---|---|---|---|---|---|---|---|---|---|---|
Set values | |||||||||||
Set temperature 1 | S | 2 | 40.0 | °C | x | x | |||||
Set temperature 2 | S | 2 | 0.0 | °C | x | x | |||||
Set speed | S | 2 | 100.0 | % | x | x | |||||
Set flow rate | S | 2 | 10.0 | L/min | x | x | |||||
Set temperature difference | S | 2 | 2.5 | K | x | x | |||||
Heating ramp | E | 3 | Value: 5.0 Status: x | K/min | x | x | |||||
Cooling ramp | E | 3 | Value: 5.0 Status: x | K/min | x | x | |||||
Safety cooling temperature | E | 3 | 70.0 | °C | x | x | |||||
Limiting temperature | E | 3 | Temperature 100: 100 Temperature 140: 140 Temperature 160: 160 Temperature 180: 180 | °C | x | x | |||||
Flow pressure limiter | E | 3 | Value: 10.0 Status: x | bar | x | x | |||||
Functions | |||||||||||
Cooling | S | 1 | x | - | x | x | |||||
Mould evacuation | S | 1 | x | - | x | x | |||||
Remote control | S | 1 | x | - | x | x | |||||
2. Set value | S | 1 | x | - | x | x | |||||
Leak stopper | S | 1 | x | - | x | x | |||||
External sensor | S | 1 | x | - | |||||||
Standby mode | S | 1 | x | - | x | x | |||||
Timer | S | 1 | x | - | x | ||||||
Display | |||||||||||
Actual values | |||||||||||
Set temperature (current) | S | - | - | °C | |||||||
Main line | S | - | - | °C | |||||||
Return line | S | - | - | °C | |||||||
External | S | - | - | °C | |||||||
Deviation actual/set | S | - | - | K | |||||||
Return/main difference | S | - | - | K | |||||||
Regulation ratio | S | - | - | % | |||||||
Flow rate | S | - | - | L/min | |||||||
Process power | S | - | - | kW | |||||||
Set value system pressure | S | - | - | bar | |||||||
System pressure | S | - | - | bar | |||||||
Main line pressure | S | - | - | bar | |||||||
Pump pressure differential | S | - | - | bar | |||||||
Return pressure | S | - | - | bar | |||||||
Pressure differential return filter | S | - | - | bar | |||||||
Operating hours | S | - | 0.0 | h | x | ||||||
Total current L1 | S | - | - | A | |||||||
Total current L2 | S | - | - | A | |||||||
Total current L3 | S | - | - | A | |||||||
Frequency converter temperature | S | - | - | °C | |||||||
Cooler temperature | S | - | - | °C | |||||||
Compensation temperature | S | - | - | °C | |||||||
External/Main line difference | S | - | - | K | |||||||
Pump output | S | - | - | % | |||||||
Pump speed | S | - | 0 | min-1 | |||||||
Pump performance | S | - | - | kW | |||||||
Pump condition | S | - | - | % | |||||||
Position cooling valve 1 | S | - | - | % | |||||||
Position cooling valve 2 | S | - | - | % | |||||||
Remaining time pump maintenance | S | 2 | 20000.0 | h | |||||||
Remaining time until heating maintenance | S | 2 | 20000.0 | h | |||||||
Remaining time cooler maintenance | S | 2 | 20000.0 | h | |||||||
Remaining time fan maintenance | S | 2 | 7000.0 | h | |||||||
Remaining time cooling water filter maintenance | S | 2 | 1000.0 | h | |||||||
Remaining time return filter maintenance | S | 2 | 1000.0 | h | |||||||
Switch cycle up to maintenance air relief valve | S | 2 | 4294967295 | - | |||||||
Switch cycle up to maintenance fill valve | S | 2 | 4294967295 | - | |||||||
Position AV 1 | S | - | - | % | |||||||
Unit information | |||||||||||
Type of unit | S | - | - | - | |||||||
Unit naming | S | 2 | Zone | - | x | x | |||||
Unit location | S | 2 | - | - | x | x | |||||
Additional equipment | S | - | - | - | |||||||
Basic type | S | - | - | - | |||||||
Maintenance | |||||||||||
Heating | S | - | 0.0 | % | |||||||
Cooling | S | - | 0.0 | % | |||||||
Pump | S | - | 0.0 | % | |||||||
Fan | S | - | 0.0 | % | |||||||
Cooling water filter | S | - | 0.0 | % | |||||||
Return line filter | S | 2 | 0.0 | % | |||||||
Filling valve | S | - | 0.0 | % | |||||||
Air relief valve | S | - | 0.0 | % | |||||||
Software licenses | S | - | - | ||||||||
Monitoring | |||||||||||
Temperature | |||||||||||
Upper set/actual deviation | S | 3 | Value: 5.0 Status: ✓ | K | x | x | |||||
Lower set/actual deviation | S | 3 | Value: 5.0 Status: ✓ | K | x | x | |||||
Return/main difference | S | 3 | Value: 10.0 Status: ✓ | K | x | x | |||||
Set/actual difference deviation | S | 3 | Value: 2.0 Status: ✓ | K | x | x | |||||
External/Main line difference | S | 3 | Value: 10.0 Status: x | K | x | x | |||||
Flow rate | |||||||||||
Maximum flow rate | S | 3 | Value: 30.0 Status: x | L/min | x | x | |||||
Minimal flow rate | S | 3 | Value: 5.0 Status: ✓ | L/min | x | x | |||||
Upper set/actual deviation | S | 3 | Value: 2.0 Status: x | L/min | x | x | |||||
Lower set/actual deviation | S | 3 | Value: 2.0 Status: ✓ | L/min | x | x | |||||
Monitoring type | S | 3 | Automatically | - | x | x | |||||
Monitoring level | S | 3 | Middle | - | x | x | |||||
Reset monitoring limits | S | 3 | 0 | - | |||||||
Startup-alarmsuppression | S | 3 | Complete | - | x | x | |||||
Alarm contact function | S | 3 | NO1 | - | x | x | |||||
Setting | |||||||||||
Remote control | |||||||||||
Network | |||||||||||
Network connection status | S | - | -- | - | |||||||
Network configuration | S | 3 | Automatically | - | x | ||||||
IP address | S | 3 | 0 | - | x | ||||||
Advanced network settings | |||||||||||
Netmask | S | 3 | 0 | - | x | ||||||
Default gateway | S | 3 | 0 | - | x | ||||||
DNS server | S | 3 | 0 | - | x | ||||||
MAC address | S | - | 00:00:00:00:00:00 | - | |||||||
IP address | S | - | 0 | - | |||||||
Remote control address | S | 3 | 1 | - | x | ||||||
Remote access via VNC | E | 3 | x | - | x | ||||||
VNC password | E | 3 | - | - | x | ||||||
Simulate Flow as a unit | E | 3 | x | - | x | ||||||
Local control system | U | 4 | x | - | x | x | |||||
Active OPC UA connections | |||||||||||
Pump control | |||||||||||
Pump operating mode | S | 3 | Normal operation | - | x | x | |||||
Pump start-up function | E | 3 | ✓ | - | x | x | |||||
Set startup speed | E | 3 | 100.0 | % | x | x | |||||
Temperature control | |||||||||||
Internal measuring point | E | 3 | Main line | - | x | x | |||||
Controller mode | E | 3 | Automatically | - | x | x | |||||
Auto-tuning | E | 3 | 0 | - | |||||||
Auto-tuning speed | E | 4 | 100.0 | % | x | x | |||||
Heating capacity limitation | E | 4 | 100 | % | x | x | |||||
Cooling capacity limitation | E | 4 | 100 | % | x | x | |||||
Control parameter P internal | E | 4 | 15 | K | x | x | |||||
Control parameter I internal | E | 4 | 60 | s | x | x | |||||
Control parameter D internal | E | 4 | Value: 1 Status: x | s | x | x | |||||
Control parameter P external | E | 4 | 150 | K | x | x | |||||
Control parameter I external | E | 4 | 180 | s | x | x | |||||
Control parameter D external | E | 4 | Value: 45 Status: ✓ | s | x | x | |||||
Filter time external sensor Controller | E | 4 | 15.0 | s | x | x | |||||
Disturbance factor | U | 4 | Value: 3.0 Status: x | - | x | x | |||||
Date/time | |||||||||||
Sync from Gate-6 | S | 3 | ✓ | - | |||||||
Date | S | 3 | YYYY-MM-DD | - | |||||||
Time | S | 3 | HH:MM | - | |||||||
Time zone | S | 3 | Europe/St_Gallen | - | |||||||
Units | |||||||||||
Temperature-unit | S | 3 | °C | - | x | x | |||||
Flow unit | S | 3 | L/min | - | x | x | |||||
Pressure unit | S | 3 | bar | - | x | x | |||||
External sensor | |||||||||||
External sensor type | S | 3 | J/Fe-CuNi | - | x | x | x | ||||
Pt 100 conductor circuit | U | 3 | 2 or 4 conductors | - | x | x | |||||
Source external sensor | S | 3 | Locally | - | x | x | x | ||||
Switch over exernal sensor | S | 3 | Automatically | - | x | x | |||||
Infrared sensor compensation | S | 3 | x | - | x | x | |||||
Emissivity | S | 3 | 1.0 | - | x | x | |||||
IR temperature compensation | S | 3 | 0.0 | °C | x | x | |||||
External sensor filter | U | 4 | 20.0 | s | x | x | |||||
Sensor external offset | U | 4 | 0.0 | K | x | ||||||
External sensor ascent corr. | U | 4 | 0.0 | % | x | ||||||
Temperature at analogue min | E | 3 | 0.0 | °C | x | x | |||||
Temperature at analogue max | E | 3 | 200.0 | °C | x | x | |||||
Energy-efficient display | |||||||||||
Electricity price | S | 3 | 0.27 | E/kWh | x | x | x | ||||
Currency | S | 3 | EUR | - | x | x | x | ||||
Operating hours/year | S | 3 | 6000 | h | x | x | x | ||||
Miscellaneous | |||||||||||
Cooling temperature | S | 3 | 35.0 | °C | x | x | |||||
Waiting time after cooling | S | 3 | Value: 5 Status: x | min | x | x | |||||
Mould evacuation time | S | 3 | ZG true: 240 | s | x | x | |||||
Mould evacuation temperature | E | 3 | 70.0 | °C | x | x | |||||
Delay mould evacuation after start | E | 3 | 90 | s | x | x | |||||
Pressure relief at shutdown | E | 3 | ✓ | - | x | x | |||||
Pressure relief duration | E | 3 | 5 | s | x | x | |||||
Rinsing time | E | 3 | 0.5 | s | x | x | |||||
Rinse interval | E | 3 | Value: 30 Status: x | min | x | x | |||||
Restart interlock | E | 3 | x | - | x | x | |||||
Show wizard at start | S | 3 | ✓ | - | x | ||||||
Quick detection hose rupture | S | 3 | ✓ | - | x | ||||||
Set speed Standby | S | 2 | 33.3 | % | x | x | |||||
Safety off via ext. contact | S | 3 | x | - | x | ||||||
Ext. contact read at mains ON | S | 3 | x | - | x | x | |||||
Switch-off delay in case overtemperature | U | 3 | Value: 30.0 Status: x | s | x | x | |||||
Post-cooling time | U | 3 | Value: 60 Status: ✓ | s | x | x | |||||
Timer | |||||||||||
Mo-Su switch on time | S | 3 | Value: 420 Status: x | - | x | x | |||||
Mo-Su switch off time | S | 3 | Value: 480 Status: x | - | x | x | |||||
Gate | S | - | - | ||||||||
Alarm logbook | S | - | - | ||||||||
Save/load | |||||||||||
Historical data | |||||||||||
Period historical data | S | 3 | 1 day | - | x | x | |||||
Export historical data | |||||||||||
Gate-6 data | |||||||||||
Importing parameterdata | |||||||||||
Importing configuration data | |||||||||||
Exporting parameterdata | |||||||||||
Exporting configuration data | |||||||||||
Start software-update | E | 4 | - | - | |||||||
Export analysis data | |||||||||||
Exporting parameterdata | |||||||||||
Importing parameterdata | |||||||||||
Exporting configuration data | |||||||||||
Importing configuration data | |||||||||||
Exporting a quality test | |||||||||||
Help system | S | - | - | ||||||||
Setup-wizard | S | - | - | ||||||||
Setup-wizard | |||||||||||
Setup-wizard Profile Page | |||||||||||
Setup-wizard Location page | |||||||||||
Setup-wizard Network page | |||||||||||
Setup-wizard Done page | |||||||||||
Service | |||||||||||
Configuration | U | - | - | ||||||||
Frequency converter | |||||||||||
Reset complete | U | 4 | 0 | - | |||||||
Reset History DB | U | 4 | 0 | - | |||||||
Parameter | |||||||||||
Cooling | |||||||||||
Stepper motors | |||||||||||
Step motor 1 | U | 3 | 0 | % | |||||||
Step motor 2 | U | 3 | 80 | % | |||||||
Step motor 3 | U | 3 | 100 | % | |||||||
Set cooler | U | 3 | 60.0 | °C | x | x | |||||
Cooling ramp 2 | U | 4 | Value: 50.0 Status: ✓ | K/min | x | x | |||||
Temperature cooling ramp 2 | U | 4 | 140.0 | °C | x | x | |||||
Pressure control | |||||||||||
Pressure control mode | U | 3 | Automatically | - | x | x | |||||
Set pressure value (manual) | U | 3 | 1.0 | bar | x | x | |||||
Air relief pressure | U | 3 | 1.5 | bar | x | x | |||||
Absolute pressure set value correction | U | 3 | 0.5 | bar | x | x | |||||
Pressure setpoint correction percentage | U | 3 | 10 | % | x | x | |||||
Pressure control hysteresis | U | 3 | 0.3 | bar | x | x | |||||
Pressure control dead band | U | 3 | 1.7 | bar | x | x | |||||
Filling pump control | U | 3 | Manual | - | x | x | |||||
Filling pump switch-on pressure | U | 3 | 1.0 | bar | x | x | |||||
Filling pump switch-on pressure (current) | U | - | - | bar | |||||||
Set pressure value max. | U | 3 | 16.0 | bar | x | x | |||||
Set pressure value min. | U | 3 | 0.7 | bar | x | x | |||||
Cooling regulation ratio limit refill | U | 4 | Cooling type A2: 100 Cooling type E2: 50 | % | x | x | |||||
Filling/air relief | |||||||||||
Max. operating time filling pump | U | 3 | 120 | s | x | x | |||||
Pausenfaktor filling pump | U | 3 | 25.0 | % | x | x | |||||
Maximum fill time | U | 3 | 30 | s | x | x | |||||
Leakage monitoring | U | 3 | Value: 300 Status: ✓ | ml/h | x | x | |||||
Leakage monitoring reset time | U | 3 | 60 | min | x | x | |||||
Venting efficiency | U | 3 | 0.5 | - | x | x | |||||
Number air relief | U | 3 | 25 | - | x | x | |||||
Fill monitoring factor | U | 3 | 5 | - | x | x | |||||
Reaction in case of incorrect measurements | U | 3 | Sinks | - | x | x | |||||
Maintenance | |||||||||||
Heating | |||||||||||
Factor heater overtemp. | U | 4 | 10000 | - | x | x | |||||
Factor heater time | U | 4 | 20000 | h | x | x | |||||
Cooling | |||||||||||
Factor cooler time | U | 4 | 20000 | h | x | x | |||||
Factor cooler overtemp. | U | 4 | 2000 | K/% | x | x | |||||
Pump | |||||||||||
Factor pump time | U | 4 | 20000 | h | x | x | |||||
Pump condition Tolerance | U | 4 | 40 | % | x | x | |||||
Minimum pump condition | U | 4 | 60 | % | x | x | |||||
Fan | |||||||||||
Fan duration factor | U | 4 | 7000 | h | x | x | |||||
Fan limit value factor | U | 4 | 30 | K/% | x | x | |||||
CPU GIF temperature limit value | U | 4 | 80.0 | °C | x | x | |||||
Cooling water filter | |||||||||||
Factor cooling water filter | U | 4 | 1000 | h | x | x | |||||
Return line filter | |||||||||||
Limit value factor return filter | U | 4 | 1000 | h | x | x | |||||
Reset maintenance filter limit value | U | 4 | 5.0 | % | x | x | |||||
Valves | |||||||||||
Factor switching cycles valves | U | 4 | 10000000 | - | x | x | |||||
Maintenance warning threshold | U | 4 | 95 | % | x | x | |||||
Command | |||||||||||
Wait reset operation release | U | 3 | Value: 300 Status: ✓ | s | x | x | |||||
Reset time menu | U | 3 | Value: 300 Status: ✓ | s | x | x | |||||
Screensaver | U | 3 | x | - | x | x | |||||
Heat transfer medium | |||||||||||
Density (water) | U | 4 | 1.0 | kg/L | x | x | |||||
Density coefficient (water) | U | 4 | -0.5 | g/L K | x | x | |||||
Thermal capacity (water) | U | 4 | 4200 | J/kg K | x | x | |||||
Coeff. thermal capacity (water) | U | 4 | 2.0 | J/kg K2 | x | x | |||||
Density (oil) | U | 4 | 1.059 | kg/L | x | x | |||||
Density coefficient (oil) | U | 4 | -0.716 | g/L K | x | x | |||||
Thermal capacity (oil) | U | 4 | 2000 | J/kg K | x | x | |||||
Coeff. thermal capacity (oil) | U | 4 | 4.0 | J/kg K2 | x | x | |||||
Leak stopper | |||||||||||
Temperature limit leak stopper | U | 3 | 70.0 | °C | x | x | |||||
Calibrating | |||||||||||
Pressure | |||||||||||
Pressure sensor filter | U | 4 | 2.0 | s | x | x | |||||
Pressure sensor 1 offset | U | 4 | 0.0 | bar | x | ||||||
Pressure sensor 1 ascent corr. | U | 4 | 0.0 | % | x | ||||||
Pressure sensor 2 offset | U | 4 | 0.0 | bar | x | ||||||
Pressure sensor 2 ascent corr. | U | 4 | 0.0 | % | x | ||||||
Pressure sensor 3 offset | U | 4 | 0.0 | bar | x | ||||||
Pressure sensor 3 slope correction | U | 4 | 0.0 | % | x | ||||||
Temperature | |||||||||||
Main/return line sensor filter | U | 4 | 20.0 | s | x | x | |||||
Main line sensor offset | U | 4 | 0.0 | K | x | ||||||
Main line sensor ascent corr | U | 4 | 0.0 | % | x | ||||||
Return line sensor offset | U | 4 | 0.0 | K | x | ||||||
Return line sensor ascent corr. | U | 4 | 0.0 | % | x | ||||||
Cooler sensor filter | U | 4 | 20.0 | s | x | x | |||||
Cooler sensor offset | U | 4 | 0.0 | K | x | ||||||
Cooler sensor ascent corr. | U | 4 | 0.0 | % | x | ||||||
Tank sensor filter | U | 4 | 20.0 | s | x | x | |||||
Tank sensor offset | U | 4 | 0.0 | K | x | ||||||
Tank sensor incline correct | U | 4 | 0.0 | % | x | ||||||
1 offset backup sensor | U | 4 | 0.0 | K | x | ||||||
Reserve sensor with 1 slope correct | U | 4 | 0.0 | % | x | ||||||
2 offset backup sensor | U | 4 | 0.0 | K | x | ||||||
Reserve sensor with 2 incline corr. | U | 4 | 0.0 | % | x | ||||||
Flow rate | |||||||||||
Flow rate filter (internal) | U | 4 | 10.0 | s | x | x | |||||
Flow rate offset (internal) | U | 4 | 0.0 | L/min | x | ||||||
Flow rate ascent corr. (internal) | U | 4 | 0.0 | % | x | ||||||
Maintenance | |||||||||||
Actual values | |||||||||||
Temperature CPU USR-61 | U | - | 0.0 | °C | |||||||
Temperature CPU GIF-61 | U | - | 0.0 | °C | |||||||
Temperature CPU ZSM-61 | U | - | 0.0 | °C | |||||||
Operating mode (current) | U | - | Undefined | - | |||||||
Error rate flow rate filtered | U | - | - | % | |||||||
Heating current 1 L1 | U | - | - | A | |||||||
Heating current 1 L2 | U | - | - | A | |||||||
Heating current 1 L3 | U | - | - | A | |||||||
Heating current 2 L1 | U | - | - | A | |||||||
Heating current 2 L2 | U | - | - | A | |||||||
Heating current 2 L3 | U | - | - | A | |||||||
Heating current 3 L1 | U | - | - | A | |||||||
Heating current 3 L2 | U | - | - | A | |||||||
Heating current 3 L3 | U | - | - | A | |||||||
Heating current 4 L1 | U | - | - | A | |||||||
Heating current 4 L2 | U | - | - | A | |||||||
Heating current 4 L3 | U | - | - | A | |||||||
Heating voltage 1 L1 | U | - | - | V | |||||||
Heating voltage 1 L2 | U | - | - | V | |||||||
Heating voltage 1 L3 | U | - | - | V | |||||||
Heating voltage 2 L1 | U | - | - | V | |||||||
Heating voltage 2 L2 | U | - | - | V | |||||||
Heating voltage 2 L3 | U | - | - | V | |||||||
Heating voltage 3 L1 | U | - | - | V | |||||||
Heating voltage 3 L2 | U | - | - | V | |||||||
Heating voltage 3 L3 | U | - | - | V | |||||||
Heating voltage 4 L1 | U | - | - | V | |||||||
Heating voltage 4 L2 | U | - | - | V | |||||||
Heating voltage 4 L3 | U | - | - | V | |||||||
Heating capacity 1 L1 | U | - | - | kW | |||||||
Heating capacity 1 L2 | U | - | - | kW | |||||||
Heating capacity 1 L3 | U | - | - | kW | |||||||
Heating capacity 2 L1 | U | - | - | kW | |||||||
Heating capacity 2 L2 | U | - | - | kW | |||||||
Heating capacity 2 L3 | U | - | - | kW | |||||||
Heating capacity 3 L1 | U | - | - | kW | |||||||
Heating capacity 3 L2 | U | - | - | kW | |||||||
Heating capacity 3 L3 | U | - | - | kW | |||||||
Heating capacity 4 L1 | U | - | - | kW | |||||||
Heating capacity 4 L2 | U | - | - | kW | |||||||
Heating capacity 4 L3 | U | - | - | kW | |||||||
Regulation ratio should | U | 4 | 0.0 | % | |||||||
Current set value controller | U | - | - | °C | |||||||
Tank temperature | U | - | - | °C | |||||||
Regulation ratio | U | 4 | - | % | |||||||
Voltage 24 VAC | U | - | - | VAC | |||||||
Voltage 24 VDC | U | - | - | VDC | |||||||
Mains frequency (current) | U | - | - | Hz | |||||||
Voltage 12 VDC GIF-61 | U | - | - | VDC | |||||||
Voltage 5 VDC GIF-61 | U | - | - | VDC | |||||||
Voltage 3.3 VDC GIF-61 | U | - | - | VDC | |||||||
Voltage 1.8 VDC GIF-61 | U | - | - | VDC | |||||||
Return/main difference | U | - | - | K | |||||||
Heating current L1 | U | - | - | A | |||||||
Heating current L2 | U | - | - | A | |||||||
Heating current L3 | U | - | - | A | |||||||
L1 phase heating voltage | U | - | - | V | |||||||
L2 phase heating voltage | U | - | - | V | |||||||
L3 phase heating voltage | U | - | - | V | |||||||
Pump current L1 | U | - | - | A | |||||||
Pump current L2 | U | - | - | A | |||||||
Pump current L3 | U | - | - | A | |||||||
Heating capacity phase 1 | U | - | - | kW | |||||||
Heating capacity phase 2 | U | - | - | kW | |||||||
Heating capacity phase 3 | U | - | - | kW | |||||||
Phase power L1 | U | - | - | kW | |||||||
Phase power L2 | U | - | - | kW | |||||||
Phase power L3 | U | - | - | kW | |||||||
Fan speed | U | - | 0 | min-1 | |||||||
Pressure waste coefficient return line filter | U | - | - | - | |||||||
Actual value controller | U | - | - | °C | |||||||
Recognized network shape | U | - | Unknown | - | x | x | |||||
Additional equipment | |||||||||||
ZL, ZG, ZF, … | U | - | x | - | |||||||
Software information | |||||||||||
Software version | U | - | - | - | |||||||
Software version GIF | U | - | - | - | |||||||
Software version DMS | U | - | - | - | |||||||
Software version ZSM 1 | U | - | - | - | |||||||
Software version ZSM 2 | U | - | - | - | |||||||
Software version ZSM 3 | U | - | - | - | |||||||
Software version ZSM 4 | U | - | - | - | |||||||
Software version FU | U | - | - | - | |||||||
DMS software update status | U | - | -- | - | |||||||
ZSM 1 software update status | U | - | -- | - | |||||||
ZSM 2 software update status | U | - | -- | - | |||||||
ZSM 3 software update status | U | - | -- | - | |||||||
ZSM 4 software update status | U | - | -- | - | |||||||
Boot loader version DMS | U | - | - | - | |||||||
Hardware-version GIF/DMS | U | - | - | - | |||||||
Serial number GIF/DMS | U | - | - | - | |||||||
Boot loader version ZSM1 | U | - | - | - | |||||||
Hardware-version ZSM1 | U | - | - | - | |||||||
Serial number ZSM1 | U | - | - | - | |||||||
Boot loader version ZSM2 | U | - | - | - | |||||||
Hardware-version ZSM2 | U | - | - | - | |||||||
Serial number ZSM2 | U | - | - | - | |||||||
Boot loader version ZSM3 | U | - | - | - | |||||||
Hardware-version ZSM3 | U | - | - | - | |||||||
Serial number ZSM3 | U | - | - | - | |||||||
Boot loader version ZSM4 | U | - | - | - | |||||||
Hardware-version ZSM4 | U | - | - | - | |||||||
Serial number ZSM4 | U | - | - | - | |||||||
Serial number FU | U | - | - | - | |||||||
Software version Rescue | U | - | - | - | |||||||
Baseboard type | U | - | -- | - | |||||||
Baseboard version | U | - | - | - | |||||||
Baseboard serial number | U | - | - | - | |||||||
Minimal software USR | U | - | - | - | |||||||
Minimal Software GIF | U | - | - | - | |||||||
minimal software GIF (USR cache) | U | - | - | - | |||||||
Minimal Software Thermo-6 | U | - | - | - | |||||||
Minimal software network | U | - | - | - | |||||||
Operating hours | |||||||||||
Operating hours unit | U | - | 0.0 | h | x | ||||||
Operating hours pump | U | - | 0.0 | h | x | ||||||
Operating hours cooler | U | - | 0.0 | h | x | ||||||
Operating hours heater 1 | U | - | 0.0 | h | x | ||||||
Operating hours heater 2 | U | - | 0.0 | h | x | ||||||
Operating hours heater 3 | U | - | 0.0 | h | x | ||||||
Operating hours heater 4 | U | - | 0.0 | h | x | ||||||
Operating hours inverter | U | - | 0.0 | h | x | ||||||
Operating hours fan | U | - | 0.0 | h | x | ||||||
Operating hours filling pump | U | - | 0.0 | h | x | ||||||
Statistic | |||||||||||
Filling pump Switch cycle | U | - | 0 | - | x | ||||||
Air relief Switch cycle | U | - | 0 | - | x | ||||||
Filling Switch cycle | U | - | 0 | - | x | ||||||
Mould evacuation Switch cycle | U | - | 0 | - | x | ||||||
Leak stopper Switch cycle | U | - | 0 | - | x | ||||||
Pressure relief Switch cycle | U | - | 0 | - | x | ||||||
Print Switch cycle /Ventilation | U | - | 0 | - | x | ||||||
Alarm contact Switch cycle | U | - | 0 | - | x | ||||||
Main contactor Switch cycle | U | - | 0 | - | x | ||||||
Total thermostat release H1 | U | - | 0 | - | x | ||||||
Total thermostat release H2 | U | - | 0 | - | x | ||||||
Total thermostat release H3 | U | - | 0 | - | x | ||||||
Total thermostat release H4 | U | - | 0 | - | x | ||||||
Act. number. trigger. Thermost. H1 | U | - | 0 | - | x | ||||||
Act. number. trigger. Thermost. H2 | U | - | 0 | - | x | ||||||
Act. number. trigger. Thermost. H3 | U | - | 0 | - | x | ||||||
Act. number. trigger. Thermost. H4 | U | - | 0 | - | x | ||||||
Number of mould evacuation | U | - | 0 | - | x | ||||||
Status | |||||||||||
Actuator air relief detected | U | - | x | - | |||||||
Actuator filling detected | U | - | x | - | |||||||
Mold emptying actuator detected | U | - | x | - | |||||||
Actuator leak stopper detected | U | - | x | - | |||||||
Actuator press. relief ZG detected | U | - | x | - | |||||||
Actuator compr. air ZG detected | U | - | x | - | |||||||
Actuator air relief active | U | 4 | x | - | |||||||
Actuator filling active | U | 4 | x | - | |||||||
Actuator mould evacuation active | U | 4 | x | - | |||||||
Actuator leak stopper active | U | 4 | x | - | |||||||
Actuator press. relief ZG active | U | 4 | x | - | |||||||
Actuator compr. air ZG active | U | 4 | x | - | |||||||
Actuator main contactor active | U | 4 | x | - | |||||||
Actuator alarm contact active | U | 4 | x | - | |||||||
Digital input 1 active | U | - | x | - | |||||||
Digital input 2 active | U | - | x | - | |||||||
Digital input 3 active | U | - | x | - | |||||||
External sensor connected | U | - | x | - | |||||||
Level measurement available | U | - | x | - | |||||||
Horn active | U | - | x | - | |||||||
Current error code flow rate | U | - | 0 | - | |||||||
Current error code ZSM1 | U | - | 0 | - | |||||||
Current error code ZSM2 | U | - | 0 | - | |||||||
Current error code ZSM3 | U | - | 0 | - | |||||||
Current error code ZSM4 | U | - | 0 | - | |||||||
Current status code ZSM1 | U | - | 0 | - | |||||||
Current status code ZSM2 | U | - | 0 | - | |||||||
Current status code ZSM3 | U | - | 0 | - | |||||||
Current status code ZSM4 | U | - | 0 | - | |||||||
Filling pump status | U | 4 | x | - | |||||||
Effective filling pump status | U | - | x | - | |||||||
Auto-tuning active | U | - | x | - | |||||||
Reset | |||||||||||
Maintenance | |||||||||||
Reset maintenance heating | U | 4 | 0 | - | |||||||
Reset maintenance cooler | U | 4 | 0 | - | |||||||
Reset maintenance pump | U | 4 | 0 | - | |||||||
Reset maintenance fill valve | U | 4 | 0 | - | |||||||
Reset maintenance air relief valve | U | 4 | 0 | - | |||||||
Reset maintenance fan | U | 4 | 0 | - | |||||||
Reset maintenance cooling water filter | U | 4 | 0 | - | |||||||
Reset maintenance return filter | U | 4 | 0 | - | |||||||
Averages | |||||||||||
Reset average main line temp. | U | - | 0 | - | |||||||
Reset average return line temp. | U | - | 0 | - | |||||||
Reset average flow | U | - | 0 | - | |||||||
Reset custom parameter | U | 4 | 0 | - | |||||||
Reset system parameter | U | 4 | 0 | - | |||||||
Reset logbook alarms | U | 4 | 0 | - | |||||||
Reset historical data | U | 4 | 0 | - | |||||||
Reset user logbook | U | 4 | 0 | - | |||||||
Averages | |||||||||||
Average main line temp. | U | - | 0.0 | °C | |||||||
Average return line temp. | U | - | 0.0 | °C | |||||||
Average flow | U | - | 0.0 | L/min | |||||||
Simulation | |||||||||||
Simulation operation | U | 3 | x | - | |||||||
External process power | U | 4 | 1.0 | kW | x | x | |||||
Activate alarm | U | 4 | Alarmlist | - | |||||||
Disable alarm | U | 4 | Alarmlist | - | |||||||
Number of DFM-51 | U | 4 | 0 | - | x | ||||||
Test | |||||||||||
Service mode | U | 3 | x | - | |||||||
Position cooling valve 1 (Nom) | U | 4 | - | % | |||||||
Position cooling valve 2 (Nom) | U | 4 | - | % | |||||||
Pump status | U | 4 | x | - | |||||||
Pump direction | U | 4 | Forwards | - | |||||||
Position AV 1 (target) | U | 4 | - | % | |||||||
Flow | |||||||||||
Overview Flow | |||||||||||
Flow A-D | |||||||||||
Settings | |||||||||||
DFM | |||||||||||
Configuration | |||||||||||
Calibrating | |||||||||||
Reset | |||||||||||
Labelling Flow | S | 3 | -- | - | |||||||
Address range Flow | S | 3 | -- | - | |||||||
Delete flow | S | 3 | 0 | - | |||||||
Number measuring circuits Flow | U | 4 | - | - | |||||||
Actual values | |||||||||||
Main line A1..D16 | S | - | °C | ||||||||
Return line A1..D16 | S | - | °C | ||||||||
Return/main difference A1..D16 | S | - | K | ||||||||
Flow rate A1..D16 | S | - | L/min | ||||||||
Monitoring | |||||||||||
Temperature | |||||||||||
Return/main difference A1..D16 | S | 0 | - | K | |||||||
Flow rate | |||||||||||
Maximum flow rate A1..D16 | S | 0 | - | L/min | |||||||
Minimal flow rate A1..D16 | S | 0 | - | L/min | |||||||
Reset | |||||||||||
Reset complete | U | 4 | 0 | - | |||||||
Reset system parameter | U | 4 | 0 | - | |||||||
Measuring circuit active A1..D16 | S | 3 | x | - | x | ||||||
User logbook | S | - | - | ||||||||
Quality check | |||||||||||
Tolerances | |||||||||||
Tolerance flow rate measurement | U | 4 | 10.0 | % | x | x | |||||
Tolerance pressure measurement | U | 4 | 0.1 | bar | x | x | |||||
Tolerance pressure gauge | U | 4 | 0.5 | bar | x | x | |||||
Internal temperature measurement Tolerance | U | 4 | 1.0 | K | x | x | |||||
External temperature measurement Tolerance | U | 4 | 6.0 | K | x | x | |||||
Tolerance external sensor Pt 100 | U | 4 | 3.0 | K | x | x | |||||
Tolerance pump pressure | U | 4 | 30 | % | x | x | |||||
Maximum heat-up time | U | 4 | 10 | min | x | x | |||||
Maximum cooling time | U | 4 | 10 | min | x | x | |||||
Start quality test | U | 4 | 0 | - | |||||||
Test temperature 3 | U | 4 | Value: 60.0 Status: x | °C | x | x | |||||
Test temperature 4 | U | 4 | Value: 80.0 Status: x | °C | x | x | |||||
Apply calibrating | U | 4 | 0 | - | |||||||
Profile | |||||||||||
Language | S | 0 | English | - | x | ||||||
User profile | S | 4 | Standard | - | x | x | |||||
Change code | S | 4 | 1234 | - | x | x | |||||
Operating release | S | 4 | 2 | - | x | x | |||||
Horn volume | E | 3 | 10 | - | x | x | |||||
Screen brightness | S | 3 | 65 | % | x | x |