qualification

Qualified personnel

Material

Pos.DesignationArticleNumber
01Ground light SPL-61T286041

Required material

  • Torx screwdriver size 20
  • Cleaning/solvent to remove adhesive residues

procedure

ATTENTION!

Working with the unit requires knowledge of the safety instructions and quick guide. That's why:

Read the safety instructions and quick guide carefully before starting any work. The basic requirement for safe work is compliance with all safety instructions and careful action by qualified qualified personnel to prevent accidents involving personal injury and property damage.

Cool down and turn off the unit

  • On the basic screen, tap the function button ().
    → The unit cools down until the temperature is lower than the cooling temperature. The unit then switches off.

  • Switch off the main switch (QS 1), pull out the mains-connector.

Open front

  • Open the front door and fold the front completely down by loosening the two Torx screws.

Replace SPL-61 (A 18) ground light

  • For BG61 only: Release and remove the cable cover (1) by pressing the mounting clips together.
  • Gently pull the cable harness out of the cable duct and push it aside so that the SPL-61 board is accessible.
  • Unplug the X922 connector on the SPL-61 board.
  • The SPL-61 board is connected to the housing with a double-sided adhesive tape. Press the circuit board into the housing from the outside to the inside until the adhesive is released.
  • Remove the SPL-61 board from the housing.
  • Remove adhesive residues from the surface of the housing, if necessary with a suitable cleaning/solvent.
    • Make sure the adhesive surface is clean and free of grease.
  • Glue the new SPL-61 circuit board into the appropriate housing recess.
  • Plug the X922 connector back into the SPL-61 board.
  • Put the cable harness back into the cable duct.
  • For BG61 only: Insert the cable cover (1) into the appropriate recesses.
    • Make sure the mounting clips are snapped in place.

Close front

  • Fold up the front and secure 2x Torx screws.

Check function

  • Connect the mains-connector and switch on the main switch (QS 1).
    • Check whether the ground light flashes with during the start process.
  • Switch on the unit using the I/O button (), check the tightness and function of the unit.

switch off unit

  • Switch off the unit using the I/O button ().
    → The unit switches off and, if necessary, is cooled and depressurised.
  • Switch off the main switch (QS 1).